Bulg. J. Phys. vol.27 no.S1 (2000), pp. 037-041
The Measurement of the Backscatter Factors for Photons of Low Energy
I.I. Tomljenović1, M.M. Ninković2
1Electrotechnical Faculty, University of Banjaluka, Bosnia and Herzegovina
2Institute of Nuclear Sciences "Vinča", Belgrade, Yugoslavia
go back1Electrotechnical Faculty, University of Banjaluka, Bosnia and Herzegovina
2Institute of Nuclear Sciences "Vinča", Belgrade, Yugoslavia
Abstract. Experimentally determined values of x-ray backscatter factors are presented in this paper. Measurements were made for x-rays generated at voltages between 60 kV and 300 kV and three water phantoms different in shape and size. To study the influence of the irradiation geometry on the backscatter factors, the measurements were performed for different photon beam field diameters at the phantom front face, at a fixed source-to-phantom distance of 1 m. The ionization chamber, volume of 1 cm3, was used for dose measurement. Measured results are analysed and discussed in comparison with measured and calculated values given in cited references.